Robotic-assisted intracranial aneurysm treatment: 1 year follow-up imaging and clinical outcomes
Nicole Mariantonia Cancelliere 1, Jeremy Lynch 2, Patrick Nicholson 2, Tomas Dobrocky 2, Saravana Kumar Swaminathan 2, Eef Jacobus Hendriks 2, Timo Krings 2 3, Ivan Radovanovic 3, Kaitlyn E Drake 4, Raymond Turner 4 5, John-Michael Sungur 4, Vitor M Pereira 6
Affiliations expand
- PMID: 34911735
- PMCID: PMC9685724
- DOI: 10.1136/neurintsurg-2021-017865
Free PMC article
J Neurointerv Sur doi: 10.1136/neurintsurg-2021-017865. Epub 2021 Dec 15.
Abstract
Background: The use of robotics in medicine may enable increased technical accuracy, reduced procedural time and radiation exposure, and remote completion of procedures. We have previously described the first-in-human, robotic-assisted cerebral aneurysm treatment using the CorPath GRX Robotic System. In this report we discuss our early experiences and outcomes using this robotic device for endovascular treatment of intracranial aneurysms using stent-assisted coil embolization and flow diversion.
Methods: The patient and disease characteristics, procedural details, and follow-up imaging and clinical outcomes of consecutive patients undergoing robotically-assisted intracranial aneurysm embolization between November 2019 and February 2020 are presented.
Results: Six patients underwent robotically-assisted embolization of intracranial aneurysms. Four of the patients were treated with a neck-bridging stent (with or without coiling) and two patients were treated with a flow-diverting stent. Two patients were treated in the subacute period of subarachnoid hemorrhage and four patients were treated electively. All of the procedures could be completed robotically and there was no need for unplanned manual intervention. The technical success rate of the procedures was 100%. There was no morbidity or mortality associated with the procedures. One year follow-up imaging showed that four aneurysms were completely obliterated (Raymond-Roy Occlusion Classification (RROC) class I) and the remaining two were occluded with a residual neck (RROC class II).
Conclusions: The Corpath GRX Robotic System demonstrated a precise control over the microcatheter, wire and stent during aneurysm treatment. Robotic neuro-procedures seem to be safe and effective and demonstrate stable occlusion results in the midterm follow-up.
Keywords: aneurysm; brain; coil; stent; technology.
© Author(s) (or their employer(s)) 2022. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.
Conflict of interest statement
Competing interests: VMP is an unpaid member of the neuro-advisory board of Corindus. KD, JMS and RT work at Corindus (robotic company).
Figures
Similar articles
- Early experience with the Penumbra SMART coil in the endovascular treatment of intracranial aneurysms: Safety and efficacy.Stapleton CJ, Torok CM, Patel AB.Interv Neuroradiol. 2016 Dec;22(6):654-658. doi: 10.1177/1591019916663479. Epub 2016 Sep 8.PMID: 27609753 Free PMC article.
- Endovascular treatment of complex intracranial aneurysms using Acandis Acclino stents.Brassel F, Grieb D, Meila D, Schlunz-Hendann M, Greling B, Melber K.J Neurointerv Surg. 2017 Sep;9(9):854-859. doi: 10.1136/neurintsurg-2016-012571. Epub 2016 Aug 24.PMID: 27557944
- Transcirculation Approach for Endovascular Embolization of Intracranial Aneurysms, Arteriovenous Malformations, and Dural Fistulas: A Multicenter Study.Roa JA, Ortega-Gutierrez S, Martinez-Galdamez M, Maud A, Dabus G, Pazour A, Dandapat S, Arteaga MS, Fernandez JG, Paez-Granda D, Kalousek V, Pons RB, Mowla A, Duckwiler G, Szeder V, Jabbour P, Hasan DM, Samaniego EA.World Neurosurg. 2020 Feb;134:e1015-e1027. doi: 10.1016/j.wneu.2019.11.078. Epub 2019 Nov 20.PMID: 31759150
- Predictors of success following endovascular retreatment of intracranial aneurysms.Mascitelli JR, Oermann EK, Mocco J, Fifi JT, Paramasivam S, Stapleton CJ, Patel AB.Interv Neuroradiol. 2015 Aug;21(4):426-32. doi: 10.1177/1591019915590070. Epub 2015 Jun 19.PMID: 26092439 Free PMC article.
- Reconstructive endovascular treatment of basilar artery fenestration aneurysms: A multi-centre experience and literature review.Styczen H, Fischer S, Gawlitza M, Meyer L, Goertz L, Maurer C, Alexandrou M, Khanafer A, Lobsien D, Deuschl C, Klisch J, Kabbasch C, Fiehler J, Berlis A, Papanagiotou P, Henkes H, Maus V.Neuroradiol J. 2022 Jun;35(3):319-328. doi: 10.1177/19714009211042877. Epub 2021 Sep 3.PMID: 34476993 Review.
References
- Lechky O. Worlds ?rst surgical robot in BC. The Medical Post 1985;21:92–3.